PRESSURE TRANSDUCER CALIBRATION Final 1

Running head: PRESSURE TRANSDUCER CALIBRATION 1
Pressure Transducer Calibration
Name
Institutional Affiliation
PRESSURE TRANSDUCER CALIBRATION 2
Contents
Objectives ..................................................................................................................................................... 3
Nomenclature ................................................................................................................................................ 3
Theory ........................................................................................................................................................... 3
Equipment ..................................................................................................................................................... 4
Measurements ............................................................................................................................................... 4
Calculations................................................................................................................................................... 5
Discussion ..................................................................................................................................................... 6
Conclusion .................................................................................................................................................... 9
Appendix .................................................................................................................................................... 10
PRESSURE TRANSDUCER CALIBRATION 3
Pressure Transducer Calibration
Objectives
The cardinal aim of this experiment was to calibrate the provided electronic transducer by
utilizing LabVIEW as well as ascertain its accuracy and working range.
Nomenclature
The experiment was concerned with calibrating a pressure transducer and using it to
measure pressure. Since voltage output was employed in the calibration process, there were only
two variables encountered in this experiment, namely pressure and voltagewhose dimensions
were (PSI) and volts (V) respectively.
Theory
Pressure measurement can be achieved through various methods. Firstly, simple pressure
gauges that are found along flow loops are examples of pressure quantification devices that
employ simple mechanical principles. However, there exist other more sophisticated tools, such
as pressure transducers, which are capable of converting mechanical pressure mechanisms to an
electrical signal. Studies indicate that pressure transducersare classified into two broad
categories, whereby the first group exhibits piezoelectric effects, while the other utilizes
aresistive strain gauge. The former kind consists of a piezoelectric material that deforms when
pressure is exerted on its surface. During the deformation process and electric potential is
generated, which is proportional to the applied stress. On the other hand, a fine wire that is
capable of deforming when subjected to pressure is used in the manufacturing of resistive strain
gauge transducers. When this wire deforms, its resistance varies depending on the intensity of the
deformation. The change in resistance is measured and converted to an electric signal to be
quantified accurately.
PRESSURE TRANSDUCER CALIBRATION 4
In this experiment, the LabVIEW software was used tocalibrate a pressure transducer.
After that, the provided VI device was modified according to the lab manual to measure pressure
obtained from the pressure tester. The set-up of the VI device was as illustrated in the figure
below. It was hypothesized that the resulting voltage would rise proportionally with an increase
in pressure and that the VI pressure readings would conform to pressure gauge readings.
Figure 1: The VI device set-up.
Equipment
Some of the equipment used to accomplish this experiment included DAQ, LabVIEW, a
pressure tester with an analog pressure gauge, an attached electronic transducer, and 12 volts DC
power supply.
Measurements
The table presented below illustrates the experimental data obtained during this
experiment.
PRESSURE TRANSDUCER CALIBRATION 5
Table 1
Pressure calibration data
P on Analog
Gauge
Measured voltage (V)
Trial 1
Trial 2
Trial 3
average
0
0.479136
0.479288
0.482497
0.480307
20
0.686099
0.589512
0.590836
0.588816
40
0.698157
0.702843
0.697851
0.699617
60
0.814951
0.815358
0.820044
0.816784
80
0.930421
0.928485
0.929504
0.92947
100
1.0408
1.04396
1.04538
0.04338
Similarly, the experimental data obtained while measuring pressure using VI was recorded in
Table 2.
Table 2
Pressure measurements:
Measured pressure (PSI)
Trial 1
Trial 2
Trial 3
average
% Deviation
10.1886
10.1886
10.1796
10.1856
1.856
29.9766
29.8864
29.3185
29.72717
0.909
49.8547
49.9718
49.8727
49.89973
0.201
69.7147
69.9491
70.5261
70.0633
0.090
90.3951
89.9083
89.9895
90.09763
0.108
Calculations
It is worth mentioning that the average measured voltage results were obtained by
summing up the voltages of the three trials under the same pressure as shown below.
For 20 PSI,
PRESSURE TRANSDUCER CALIBRATION 6
Av voltage =

= 0.588816V
The average values of all the other measured voltages were obtained in the same manner and
recorded in table 1. Similarly, the average measured pressure was obtained by summing resulting
pressure for all the trials as indicated below.
For 10PSI,
Av measured pressure =

= 10.1856PSI
The same approach was utilized to obtain average values of all the other measured pressures, and
the resulting values were recorded in Table 2. However, the measured pressure differed from the
expected values slightly. The percentage deviation of measured pressure from expected pressure
was evaluated as shown below.
For 30 PSI,
% deviation =



% deviation =


  = 1.856%
It is important to point out that the same approach was utilized to obtain the percentage deviation
of all measured pressure from the expected pressure, and the resulting data was recorded in Table
2.
Discussion
Based on the experimental results in Table 1, it was observed that as the applied pressure
increased, the resulting voltage heightened proportionally. In particular, approximately
0.480307V and 1.04338V were measured when 0 and 100PSI pressure was applied to the
PRESSURE TRANSDUCER CALIBRATION 7
transducer respectively. Notably, the proportional increase conformed to the theoretical
operational principles of a pressure transducer. In particular, when the pressure was exerted, the
tension on the transducer’s diaphragm increased, thus leading to higher voltage generation.
It should be understood that when stress on the diaphragm is translated using electrical
components, an electrical signal is generated to measure voltage. One salient point to note is that
as the pressure increased, the tension on the diaphragm rose, hence resulting in a higher voltage
reading. The resulting voltage presented in Table 1 was used to plot the graph in the figure
below.
Figure 2: The graph of calibrated pressure against voltage.
The graph shows a straight line conforming to the observations made above that as the
pressure increased, the voltage reading also rose systematically. Furthermore, the R
2
value of
0.9999 illustrates that the regression line had the best fit. Nevertheless, the experimental results
in Table 2 had slight deviation from the expected outputs. In this case, the ainticipated outcomes
pertained to the set pressure on the analog pressure gauge and the resulting pressure obtained
y = 176.99x - 84.467
R² = 0.9999
0
20
40
60
80
100
120
0 0.2 0.4 0.6 0.8 1 1.2
Pressure (PSI)
Voltage (V)
Pressure Calibration Data
PRESSURE TRANSDUCER CALIBRATION 8
from the VI display. Specifically, it was observed that the VI screen recorded an average
pressure of 70.0633 PSI when the analog pressure gauge was set to 70 PSI. Notably, this presents
an insignificant deviation of approximately 0.0633 PSI. The experimental results shown in Table
2 were used plot a graph of expected pressure against measured pressure as indicated in the
figure provided below.
Figure 3: The graph of set pressure against measured pressure.
From the graph, it is noticeable that the measured pressure conformed to the expected
pressure. To be specific, the gradient of the line was approximately 1, illustrating that all the
points had the same value in the x- and y-axes. Moreover, the regression line had an R
2
value of
1, which shows that all the points were along the line of best fit. Most importantly, the evaluation
of the percentage deviation shows that there were insignificant discrepancies between the
measured and expected results. For instance, the highest percentagedifference recorded during
this experiment was 1.856%, which was below an accepted value of 2.5%.
y = 0.9992x + 0.0469
R² = 1
0
10
20
30
40
50
60
70
80
90
100
0 20 40 60 80 100
Pressure (PSI)
Pressure (psi)
Pressure Calibration Data
PRESSURE TRANSDUCER CALIBRATION 9
Conclusion
From the experimental results presented above, it can be observed that the aim of this
experiment of calibrating a pressure transducer and further ascertaining its accuracy was attained
successfully. Notably, it was noted that as the applied pressure increased, the resulting voltage
also heightened. For example, voltage values of 0.588816V and 0.92947V were recorded when
the pressure tester was set to 20 and 80 PSI respectively. Moreover, the plot of pressure against
resulting voltage had a regression line with an R
2
value of 0.999, thus showing that all the points
were along the line of best fit. Therefore, it can be said that the experiment was conducted
precisely. In the second section of this lab, it is observed that the measured pressure has a
negligible deviation from the expected pressure. For example, a pressure of 90.09763 PSI was
recorded when the pressure tester was set to 90 PSI. Notably, the highest percentage deviation
recorded in this experiment was 1.856%, which illustrates that the lab was conducted accurately.
Moreover, the plot of measured pressure had a regression line with an R
2
of 1. Hence, all the
points fit on the line of best-fit link squarely, depicting that the experiment was also conducted
precisely. However, the slight deviation recorded could indicate that some errors were
encountered during this experiment. Some of the mistakes that might have been encountered
include the use of a faulty VI device that may have caused erratic voltage reading, which may
have lowered the precision of the results in the first part of the lab. Similarly, the pressure gauge
might have had some problems, thus leading to inaccurate expected pressure readings.
Nevertheless, the mistakes did not hamper the results significantly.
PRESSURE TRANSDUCER CALIBRATION 10
Appendix
Figure 4: The experimental set-up used to calibrate pressure transducer.
Figure 5: The schematic block diagram of VI set-up.

Place new order. It's free, fast and safe

-+
550 words

Our customers say

Customer Avatar
Jeff Curtis
USA, Student

"I'm fully satisfied with the essay I've just received. When I read it, I felt like it was exactly what I wanted to say, but couldn’t find the necessary words. Thank you!"

Customer Avatar
Ian McGregor
UK, Student

"I don’t know what I would do without your assistance! With your help, I met my deadline just in time and the work was very professional. I will be back in several days with another assignment!"

Customer Avatar
Shannon Williams
Canada, Student

"It was the perfect experience! I enjoyed working with my writer, he delivered my work on time and followed all the guidelines about the referencing and contents."

  • 5-paragraph Essay
  • Admission Essay
  • Annotated Bibliography
  • Argumentative Essay
  • Article Review
  • Assignment
  • Biography
  • Book/Movie Review
  • Business Plan
  • Case Study
  • Cause and Effect Essay
  • Classification Essay
  • Comparison Essay
  • Coursework
  • Creative Writing
  • Critical Thinking/Review
  • Deductive Essay
  • Definition Essay
  • Essay (Any Type)
  • Exploratory Essay
  • Expository Essay
  • Informal Essay
  • Literature Essay
  • Multiple Choice Question
  • Narrative Essay
  • Personal Essay
  • Persuasive Essay
  • Powerpoint Presentation
  • Reflective Writing
  • Research Essay
  • Response Essay
  • Scholarship Essay
  • Term Paper
We use cookies to provide you with the best possible experience. By using this website you are accepting the use of cookies mentioned in our Privacy Policy.