PRESSURE TRANSDUCER CALIBRATION 9
Conclusion
From the experimental results presented above, it can be observed that the aim of this
experiment of calibrating a pressure transducer and further ascertaining its accuracy was attained
successfully. Notably, it was noted that as the applied pressure increased, the resulting voltage
also heightened. For example, voltage values of 0.588816V and 0.92947V were recorded when
the pressure tester was set to 20 and 80 PSI respectively. Moreover, the plot of pressure against
resulting voltage had a regression line with an R
2
value of 0.999, thus showing that all the points
were along the line of best fit. Therefore, it can be said that the experiment was conducted
precisely. In the second section of this lab, it is observed that the measured pressure has a
negligible deviation from the expected pressure. For example, a pressure of 90.09763 PSI was
recorded when the pressure tester was set to 90 PSI. Notably, the highest percentage deviation
recorded in this experiment was 1.856%, which illustrates that the lab was conducted accurately.
Moreover, the plot of measured pressure had a regression line with an R
2
of 1. Hence, all the
points fit on the line of best-fit link squarely, depicting that the experiment was also conducted
precisely. However, the slight deviation recorded could indicate that some errors were
encountered during this experiment. Some of the mistakes that might have been encountered
include the use of a faulty VI device that may have caused erratic voltage reading, which may
have lowered the precision of the results in the first part of the lab. Similarly, the pressure gauge
might have had some problems, thus leading to inaccurate expected pressure readings.
Nevertheless, the mistakes did not hamper the results significantly.