7
NA = numerical aperture of objective used in microscope.
Combining both equations and including the conversion from filar units to millimeters in the
table, the equation becomes:
KHN = P/ (LK=7 ƛ /2NA)
2
C
p
Where:
L = the length of the long diagonal in filar units as measured with the light microscope
K = the fraction of millimeter (calibration constant) represented by a filar unit.
5. Discussion of the Results
The results are consistent. There was a variation of +_0.0002 when the specimen was
measured more than twice. These variations were not excessive in that it could affect the
experiment. When the results vs. theory are compared, it corresponds to each other, in that the
experiment reflects the theory. Although there were differences between the two, the differences
were not enough to discredit either. One of the sources of error in the experiment was placement
of the specimen. Additionally, the inclination of the microscope was a source of error. Both
sources of errors were corrected during the experiment.
6. Conclusions and Recommendations
In conclusion, the experiment was successful and the Knoop hardness of glass was
identified. The experiment had a few errors but all were corrected before arriving at the final
value. It is recommended that a person taking the experiment should be very vigilant to avoid
unnecessary errors which might lead to a wrong final value.